Spirometry data in the EHR: two rules that now contradict each other
For practices doing clinical spirometry, occupational surveillance, or both on the same equipment.

Since 2023 the clinical standard has been race-neutral interpretation using average reference equations. One OSHA standard still requires race-specific values and a 0.88 multiplier. Spirometry data in the EHR therefore has to record which reference equation produced each result, per test.
A practice doing both clinical and occupational spirometry is subject to two contradictory requirements on the same machine.
In April 2023 the American Thoracic Society published an official statement recommending that laboratories adopt a race-neutral approach to test interpretation by reporting and interpreting results using average reference equations. The rationale is stated plainly in the document: the superficial appearance of race should not be used to infer biological characteristics.
Meanwhile the OSHA cotton dust standard still directs that values be compared to appropriate race and ethnicity-specific lower limit of normal values, and instructs that reference values for Asian-Americans be obtained by multiplying the Caucasian predicted values by 0.88. That regulation is current.
This is not a transitional wrinkle that resolves itself. It is two live requirements pointing in opposite directions, and a single device-level setting cannot satisfy both.
Reference equation selection is now a decision, not a default
The society statement names its recommended option: the Global Lung Function Initiative average equation, published as GLI Global, is a recommended race-neutral average reference equation. A majority of workshop participants recommended a race-neutral average equation instead of race-specific ones.
The statement is candid about what that equation is. It describes GLI Global as a race composite and not truly race agnostic — race-neutral in application rather than in construction. The peer-reviewed paper behind it notes the equations had a similar fit to the earlier other equations with wider limits of normal, which is the operationally important phrase: switching changes who falls below the lower limit of normal, in both directions.
None of this is regulation. It is a professional society statement and should be described as one. The regulation is the part that points the other way.
Occupational spirometry records live under different rules
Three separate regimes govern occupational testing, and they differ on who may test and how long records are kept. The silica standard requires a pulmonary function test administered by a spirometry technician with a current certificate from a NIOSH-approved course, within an examination performed by a licensed health care professional.
The coal programme rule requires that each person administering tests successfully complete a NIOSH-approved training course, and it incorporates a 2005 society standard by reference — so that federal rule is pinned to a document two revisions behind the current technical statement.
That retention window is the reason the reference equation has to travel with the result. A test read in 2026 may be re-read in 2050 by someone adjudicating a claim, and a value with no recorded equation is a number without a denominator.
Bronchodilator responsiveness criteria changed, and the two are coupled
The 2022 interpretive standard moved the responsiveness threshold to a change of more than 10 per cent relative to the predicted value for forced expiratory volume in one second or forced vital capacity. The 2005 criterion was a change of at least 12 per cent and at least 200 millilitres of the initial value.
Read what moved. The denominator went from baseline to predicted. A template still computing twelve per cent of baseline is calculating the 2005 rule under a 2026 date stamp.
And because predicted values come from the reference equation, the two issues are not separate. Changing the equation changes the predicted value, which changes the responsiveness classification. A build that treats them as unrelated settings will produce results that disagree with the device. Practices running pulmonology EHR and IT should check both settings together rather than in sequence.

Does your system record which reference equation produced the result?
We will look at how spirometry results land in your record and whether the fields a reviewer needs in 2040 are being captured today.
Test quality grading belongs in a field
The 2019 technical statement sets acceptability and repeatability criteria that most systems never capture. Back-extrapolated volume must be no more than five per cent of forced vital capacity or 0.100 litres, whichever is greater. For patients over six, the difference between the two largest values must be no more than 0.150 litres, for both measures. A minimum of three acceptable measurements is expected.
The statement also distinguishes an acceptable manoeuvre from a merely usable one, and introduced a grading system so an interpreter knows which produced the reported values. That grade is a structured field, and in most systems it does not exist — which means the person reading the report cannot tell whether the number came from three good efforts or one salvageable one.
Interpretation and report: bundled, but still required
For billing purposes the spirometry codes include the laboratory procedure and the interpretation of test results. The interpretation is not separately reportable — but the record still has to contain one for the code to be supported.
The wider documentation requirement comes from the diagnostic test rules. The test must be ordered by the physician treating the beneficiary, the ordering practitioner must maintain documentation of medical necessity in the record, and on review the billing entity must produce documentation of the order including enough information to identify and contact the ordering practitioner. Getting those fields onto the encounter form is template development work, and it sits inside our EHR solutions.
Spirometry in the record: common questions
Pick one study from the last month and see which predicted set the software applied. Write down whether that set is the race-neutral one the clinical standard now uses or the race-specific one the OSHA rule still names, and do not leave both unlabeled on the same row. The person who signs the report and the person who configures the device should look at that same study together. If they cannot say which rule the number came from, the chart is not ready for the next occupational exam.
Should we turn off race correction in our spirometry system?
Which reference equations should a clinical build use?
Did the bronchodilator responsiveness threshold change?
How long do occupational spirometry records have to be kept?
Who is allowed to perform occupational spirometry?
Can we bill separately for interpreting the test?
Does HIPAA treat occupational spirometry differently from clinical testing?
Let’s meet
Qventive supports pulmonology, allergy and occupational medicine practices across New Jersey. Thirty years healthcare IT, one team for EHR and infrastructure.
- The recommendation to adopt a race-neutral approach using average reference equations, and the description of GLI Global as a race composite, published 15 April 2023. pmc.ncbi.nlm.nih.gov
- The requirement to compare values to race and ethnicity-specific lower limit of normal values and the 0.88 multiplier, current as of 22 September 2026. ecfr.gov
- The requirement to preserve each employee medical record for the duration of employment plus thirty years. osha.gov
- Acceptability and repeatability criteria and the quality grading system, published 15 October 2019. pmc.ncbi.nlm.nih.gov
- NIOSH-approved training requirements, course length and certificate validity, last reviewed 25 February 2025. cdc.gov